Control Unit for Critical Potential Tubes
Control unit for operating the critical potential tubes; equipped with an output for saw tooth acceleration voltages; adjustable upper and lower limits of the acceleration voltage; integrated pico-ammeter amplifier for anode-current measurement. Allows recording of the acceleration voltage as a function of the anode current. A slow saw tooth voltage (approx. 6 seconds per cycle) is available with an interface or XY-recorder; a saw tooth voltage with a repetition rate of 20 Hz is available for oscilloscopic observations.The anode current, acceleration voltage as well as their upper and lower limits can be represented directly in conjunction with the digital display with an adjustable threshold (U18620).
Input: Anode current measurement via a BNC jack
Outputs:
• Tube: Saw tooth acceleration voltage of 0 to 60 V, 20 Hz
• Fast: Voltage signal of 0 to 1 V proportional to the acceleration
voltage, for oscilloscopic observations
• Slow: Voltage signal of 0 to 1 V proportional to the acceleration
voltage, for recording data with an XY-recorder or interface
• Anode current: Voltage signal of 0 to 1 V proportional to the anode current (1V/nA)
Dimensions: 170 mm x 105 mm x 45 mm
Additionally required:
U18700 Plug-in power supply
Item: U186501
GBP 214.00 (+ VAT)
GBP 214.00 (+ VAT)